<10. Other Functions>
10-14
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
Table 10.6
Initialization Items and Default Values (Humidity Analyzer)
Item
Initialization Parameter
Default setting
Equipment
selection
Type of equipment
Not initialized
Detector
ZR22
Sample gas
Wet gas
Displayed data
Display item
1st display item
Humidity
2nd display item
Current output 1
3rd display item
Current output 2
Tag name
Deleted
Trend graph
Parameter
Humidity
Sampling interval
30 seconds
Upper limit (graph)
25 vol%O
2
Lower limit (graph)
0 vol%O
2
Automatic return time
0 min.
Language
Not initialized
Calibration
data
Calibration
setting
Mode
Manual
Calibration procedure
Span - zero
Zero gas concentration
1.00 vol%O
2
Span gas concentration
21.00 vol%O
2
Output hold time
10 min., 00 sec.
(Semi_Auto)
Calibration time
10 min., 00 sec.
(Semi_Auto)
Interval
30 days, 00 hr. (Auto)
Start date
01/01/00 (Auto)
Start time
0:00 (Auto)
Blow back
Blow back
setting
Mode
No function (invalid)
(Output) hold time
10 min., 00 sec.
Blow back time
10 min., 00 sec.
Interval
30 days, 00 hr.
Start date
01/01/00
Start time
0:00
Current output
data
mA-output 1
mA-output 2
Parameter
Humidity
Min. oxygen concentration
0 vol%O
2
Max. oxygen concentration
25 vol%O
2
Output damping constant
0 sec.
Output mode
Linear
Output hold
setting
Warm-up
4 mA
Set value of Warm-up
4 mA
Maintenance
Previous value held
Set value of Maintenance
4 mA
Calibration, blow back
Previous value held
Set value of Calibration, blow back
4 mA
Error
Preset value held
Set value of Error
3.4 mA
Summary of Contents for EXAxt ZR22G
Page 15: ...Blank Page...
Page 51: ...Blank Page...
Page 77: ...Blank Page...
Page 137: ...Blank Page...
Page 171: ...Blank Page...
Page 195: ...Blank Page...
Page 197: ......
Page 199: ......
Page 201: ......
Page 203: ......
Page 205: ......
Page 207: ......
Page 211: ...Blank Page...