<10. Other Functions>
10-12
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
Item
Initialization Parameter
Default setting
Current output
data
mA-output 1
mA-output 2
Parameter
Oxygen concentration
Min. oxygen concentration
0 vol%O
2
Max. oxygen concentration
25 vol%O
2
Output damping constant
0 sec.
Output mode
Linear
Output hold
setting
Warm-up
4 mA
Set value of Warm-up
4 mA
Maintenance
Previous value held
Set value of Maintenance
4 mA
Calibration, blow back
Previous value held
Set value of Calibration, blow back
4 mA
Error
Preset value held
Set value of Error
3.4 mA
Alarm data
Alarm setting
Parameter
Oxygen concentration
Hysteresis
0.1 vol%O
2
Delayed action of alarm contact
3 seconds
Alarm set value
High-high alarm
OFF
Alarm value of High-high alarm
100 vol%O
2
High limit alarm
OFF
Alarm value of High limit alarm
100 vol%O
2
Low limit alarm
OFF
Alarm value of Low limit alarm
0 vol%O
2
Low-low alarm
OFF
Alarm value of Low-low alarm
0 vol%O
2