6-6
IM DLM5058-03EN
Item
Specifications
Window:
Triggers when the window conditions of a single trigger source are met.
The source can be selected from CH1 to CH8.
1
Triggers when the source enters or exits from the voltage span set with the upper
and lower levels. Further, triggers can be activated on the time that the source
remains inside or outside the voltage span.
The time condition can be set to None, More than, Less than, Inside, Outside, or
Timeout.
For details on the time settings and time accuracies, see “Pulse Width.”
Window OR:
Triggers on the OR of the window conditions of multiple trigger sources.
The source can be selected from CH1 to CH8.
1
Triggers when the source enters or exits from the voltage span set with the upper
and lower levels of each channel. Enter or exist can be set separately for each
source.
Interval:
Triggers on the time from the rising to the next rising edge or the time from the falling
to the next falling edge of a single trigger source.
The source can be selected from CH1 to CH8,
1
and Logic.
The time condition can be set to None, More than, Less than, Inside, Outside, or
Timeout.
For details on the time settings and time accuracies, see “Pulse Width.”
FlexRay
2
:
Triggers on a FlexRay bus signal
The source can be selected from CH1 to CH8.
1
Mode:
Frame Start, Error, ID/Data, ID OR
Bit Rate:
2.5M, 5M, 10Mbps
Bus channel: A, B
CAN
2
:
Triggers on a CAN (Controller Area Network) bus signal
The source can be selected from CH1 to CH8.
1
Mode:
SOF, Error, ID/Data, ID OR
Bit Rate: 33.3k, 83.3k, 125k, 250k, 500k, 1Mbps, or User Define
For User Define, you can set a value from 10 k to 1 Mbps in 0.1-kbps
steps.
CAN FD
2
:
Triggers on a CAN FD (CAN with Flexible Data-Rate) bus signal (ISO or non-ISO)
The source can be selected from CH1 to CH8.
1
Mode:
SOF, Error, ID/Data, ID OR, FDF, ESI
Bit Rate: Arbitration phase
250k, 500k, 1Mbps, or User Define
For User Define, you can set a value from 20 k to 1 Mbps in 0.1-kbps
steps.
Data phase
500k, 1M, 2M, 4M, 5M, 8Mbps, or User Define
For User Define, you can set a value from 250 k to 10 Mbps in 0.1-kbps
steps.
LIN
2
:
Triggers on a LIN (Local Interconnect Network) bus signal
The source can be selected from CH1 to CH8.
1
Mode:
Break Synch, Error, ID/Data, ID OR
Bit Rate: 1200, 2400, 4800, 9600, 19200bps, or User Define
For User Define, you can set a value from 1 k to 20 kbps in 0.01-kbps
steps.
CXPI
2
:
Triggers on a CXPI (Clock Extension Peripheral Interface) bus signal
The source can be selected from CH1 to CH8.
1
Mode:
SOF, Error, PTYPE, ID/Data, ID OR, Wakeup/Sleep
Bit Rate: 4800, 9600, 19200bps, or User Define
For User Define, you can set a value from 4 k to 50 kbps in 0.01-kbps
steps.
SENT
2
:
Triggers on a SENT signal (J2716 APR2016 and earlier)
The source can be selected from CH1 to CH8,
1
and Logic.
Mode:
Every Fast CH, Fast CH S&C (Status & Communication),
Fast CH Data, Every Slow CH, Slow CH ID/Data, Error
Clock period: 1 μs to 100 μs (0.01μs resolution)
1 CH1 to CH4 for DLM5034, DLM5054
2 FlexRay, CAN, CAN FD, LIN, CXPI, SENT, PSI5 Airbag, UART, I
2
C, and SPI are options.
6.2 Triggering Section