6-16
IM DLM5058-03EN
Other Features
Item
Specifications
Default setup
Resets the instrument to its factory default settings.
The following settings are not reset: date and time settings, communication interface
settings, settings stored to the internal memory, language settings, and font size of
measured values.
The Undo command can be used to revert to the previous settings.
Auto setup
Automatically sets the voltage scale, time scale, trigger, and other settings to the most
suitable values for the input signals.
The Undo command can be used to revert to the previous settings.
Serial bus auto setup
1
Auto setup can be executed on the basis of the selected serial bus type (FlexRay,
CAN, CAN FD, LIN, CXPI, SENT, PSI5 Airbag, UART, I
2
C, SPI) and trigger source. The
instrument can automatically set the bit rate, source level, and other settings and trigger on
the basis of these settings.
Calibration
Auto calibration and manual calibration are available.
System configuration
The date, time, and message language can be specified. The click sound can be turned on
and off.
Time synchronization feature
Synchronization source: Supports IEEE1588-2008 (PTP v2) (slave only)
Supports PTP packets of Layer3 (UDP/IPv4) and Layer2 (Ethernet)
Supports Ordinary Clock
Supports E2E delay correction
Supports 2-step Sync messages
Synchronized items: Instrument clock, sampling clock
Synchronization accuracy: ±200 ns (typical value
2
, 1000BASE-T without using an Ethernet
switch)
Probe compensation signal output
Transmits signals from the front-panel probe compensation signal output terminals (approx.
1-V
p-p
and approx. 1-kHz rectangular wave).
Overview
Allows you to view the instrument system status.
Additional option licenses
Computation, trigger, analysis, and other options can be purchased and added.
Self-test
Memory, accuracy, keyboard, printer, and synchronous operation feature tests are
available.
menu language
The menu language can be changed.
Help
Displays a description of the settings. Available in English and Japanese.
1 Serial bus auto setup is available on models with the serial bus option.
2 Typical values represent typical or average values. They are not strictly warranted.
6.5 Features