6-24
IM DLM5058-03EN
Item
Specifications
Immunity
Compliant standards
EN 61326-1 Table 2 (for use in industrial locations)
(Applies to the DLM5034, DLM5038, DLM5054, and DLM5058)
EN 61326-2-1
When a measurement lead or probe is connected or when the device under measurement is
connected to the instrument, the immunity requirements may no longer be met. In such cases, the
user will be required to take appropriate measures.
Influence in the immunity testing environment (criteria A)
Noise increase: 500 µV/div to 10 mV/div: ±4 div
20 mV/div to 10 V/div: ±2 div
Test conditions: 2.5 GS/s, envelope mode
Cable conditions: Same as the emission cable conditions
Test items:
1. Static discharge: EN 61000-4-2
Air discharge: ±8 kV. Contact discharge: ±4 kV. Criteria B.
2. Radiated immunity: EN 61000-4-3
80 MHz to 1 GHz, 10 V/m, 1.0 GHz to 6.0 GHz, 3 V/m, Criteria A
3. Conducted immunity: EN 61000-4-6
10 V, Criteria A
4. EFT/Burst: EN 61000-4-4
Power line: ±2 kV. Signal line: ±1 kV, Criteria B
5. Surge immunity: EN 61000-4-5
±1 kV between lines, ±2 kV common, Criteria B
6. Voltage dip and interruptions: EN 61000-4-11
1 cycle, both polarities, 100%, criteria B
Other tests, criteria C
Definitions of criteria A, B, and C
Criteria A: During testing, "influence in the immunity testing environment"
described above is met.
Criteria B: The instrument continues to function and is controllable throughout
testing. The instrument does not change operation modes, and data
changes do not persist.
Criteria C: During testing, temporary degradation of performance or loss of
functionality occurred, the correction of which required user operation
or system reset.
Environmental standards* EU RoHS Directive compliant
* For conformity to environmental regulations and/or standards other than EU, contact your nearest YOKOGAWA office
(PIM113-01Z2).
6.12 General Specifications