6-22
IM DLM5058-03EN
6.12 General Specifications
Item
Specifications
Standard operating conditions
Ambient temperature:
23 ± 5°C
Ambient humidity:
55 ± 10%RH
Supply voltage and
frequency errors:
Within ±1% of rating
Warm-up time
At least 30 minutes
Storage environment
Temperature:
–20 to 60°C
Humidity:
20 to 80% RH (no condensation)
Altitude:
3000 m or less
Operating environment
Temperature:
5 to 40°C
Humidity:
20 to 80%RH (when the printer is not in use; no condensation)
35 to 80%RH (when the printer is in use; no condensation)
Altitude:
2000 m or less
Installation
Indoors use
Recommended calibration period 1 year
Rated supply voltage
100 to 120 VAC, 220 to 240 VAC (auto switching)
Permitted supply voltage range
90 to 132 VAC, 198 to 264 VAC
Rated supply frequency
50/60 Hz
Permitted supply frequency range 48 to 63 Hz
Power fuse
Built in (not replaceable)
Maximum power consumption
290 VA (when the printer is in use)
Withstand voltage (between the
power supply and case)
1.5 kVAC for 1 minute
Insulation resistance (between the
power supply and case)
500 VDC, 10 MΩ or more
Dimensions
426 mm (W) × 266 mm (H) × 180 mm (D), with the printer cover closed, excluding protrusions
Weight
Approx. 7.3 kg (excluding options)
Instrument cooling method
Forced air cooling, outlet on the rear
Installation orientation
Horizontal or tilted using the movable leg or stand
Other installation orientations are prohibited.
Stacking prohibited.
Battery backup
The clock is backed up with the internal lithium battery.
Battery life: Approx. five years (at an ambient temperature of 25°C)
Safety standards
Compliant standards
EN 61010-1
Overvoltage Category II
1
Pollution degree 2
2
EN 61010-2-030
Measurement Category Other (O)
3
1 The overvoltage category is a value used to define the transient overvoltage condition and includes the rated impulse
withstand voltage. The overvoltage category II applies to electrical equipment that is powered through a fixed installation, such
as a switchboard.
2 Pollution Degree applies to the degree of adhesion of a solid, liquid, or gas that deteriorates withstand voltage or surface
resistivity. Pollution degree 2 applies to normal indoor atmospheres (with only non-conductive pollution).
3 The measurement category of this instrument’s signal input terminals is Other (O). Do not use it to measure the main power
supply or for Measurement Categories II, III, and IV.
Measurement category O applies to measurement of circuits that are not directly connected to a main power source. This
category applies to measurement of secondary electric circuits in equipment across a transformer. The estimated transient
overvoltage that may appear at the instrument’s signal input terminals is 1600 V.
Measurement category II applies to measurement of circuits, such as household electric appliances and portable electric
tools, that are connected to low-voltage installations.
Measurement category III applies to measurement of facility circuits, such as distribution boards and circuit breakers.
Measurement category IV applies to measurement of power source circuits, such as entrance cables to buildings and cable
systems, for low-voltage installations.