Troubleshooting, Maintenance, and Inspection
1
2
3
4
5
6
App
Index
Code Message and Corrective Action
Section
681
The data length that is necessary for the harmonics analysis function is short.
—
682
The decode cannot be displayed, because the threshold level is not appropriate.
—
683
Cannot execute the math function, because the display of source is OFF.
Chapter 6
1
684
Logic waveforms cannot be loaded into Ref.
Chapter 17
1
685
Cannot load into Ref with maximum record length.
—
686
Cannot execute during roll mode.
Stop and execute again.
3.9
687
It connects with PC Application.
—
688
Cannot execute on current record length.
—
689
Cannot execute during Preview mode.
—
691
Cannot execute after hisory search.
Reset history search and execute again.
15.2
1
692
Cannot execute, because history is not exist.
—
693
Cannot execute, when the output of the print is “Multi”.
16.6
1
695
Failed to establish a DLMsync connection.
Check that the DLMsync cable is connected properly, and execute again.
19.1
1
696
DLMsync cannot be used because the firmware versions are not matched.
Update the firmware to the latest version.
—
697
DLMsync cannot be used because the /SYN option is not installed in the connected
instrument.
Connect an instrument with the /SYN option, or purchase the option and install it.
—
698
Failed to execute DLMsync auto deskew.
The proper signal may not be applied to CH1 of both instruments, or the probe may not be
connected properly.
See the help or the Features Guide for the execution conditions of auto deskew.
Chapter 22
2
1 User’s Manual, IM DLM5058-02EN
2 Features Guide, IM DLM5058-01EN
Setting Errors
Code Message and Corrective Action
Section
800
Improper Date / Time setting.
3.5
801
Not allowed unless waveforms are shown. Display waveforms.
Chapter 1
1
802
Source waveforms do not exist. Display source waveforms.
Chapter 3
1
803
Zone waveforms do not exist.
Chapter 2
1
804
Illegal expression.
Chapter 6
1
806
Invalid bit assignment in the logic bus.
Chapter 1
1
807
Unable to enable the trigger conditions.
Set the clock source to another group or assign bits to the group.
2.6
1
808
Cannot set this parameter with maximum record length.
—
809
Cannot change the setting of the math operation, because power analysis is set.
—
810
Cannot set this parameter when A-trigger is not serial bus.
Chapter 2
1
811
This setting is necessary only in the case of ON display of Zoom1 and Zoom2.
Chapter 10
1
812
Cannot set this parameter during interleave mode.
—
813
This option is not available.
—
814
The Userdefined math option is not available.
—
815
The logic input option is not available.
—
816
This function is not supported.
—
817
The ID value cannot be set to 0.
2.17,
1
12.5
1
825
This operation cannot be executed during DLMsync.
Chapter 22
2
826
This operation cannot be executed on the sub unit during DLMsync.
Chapter 22
2
827
This operation cannot be executed on the sub unit during DLMsync. Execute it on the main
unit.
Chapter 22
2
828
This operation cannot be executed during DLMsync.
Stop the acquisition or wait for it to finish, and then execute.
Chapter 22
2
1 User’s Manual, IM DLM5058-02EN
2 Features Guide, IM DLM5058-01EN
5.2 Messages and Corrective Actions