2-15
IM DLM5058-03EN
Measurement Preparation
1
2
3
4
5
6
App
Index
Connecting a Probe Supported by the Probe Interface Connectors
• If you connect a probe supported by the probe interface connector to the instrument, the probe
type is automatically recognized, and the attenuation ratio set.* Also because power is supplied
to the probe through the probe interface, it is not necessary to connect the probe power cable to
the probe power terminals.
* For a list of compatible probes, see “Optional Accessories” on page vii.
• You can execute automatic zero adjustment on a current probe that is compatible with the probe
interface connector.
Connecting an FET Probe, Current Probe, or Differential Probe
If you are using a YOKOGAWA FET probe, current probe, or differential probe, use one of the probe
power terminals (option) on the instrument’s right side panel for the power supply.* For details on
the connection procedure, see the manual that came with the product that you want to use.
* For a list of probe, see “Optional Accessories” on page vii.
CAUTION
Do not use the probe power terminals (option) on the instrument’s right side panel for
purposes other than supplying power to an FET probe, current probe, or differential probe.
Be sure that the total current of the eight probe power terminals and the eight probe interface
terminals (±12 V each) on 8 channel models or that of the four probe power terminals and
the four probe interface terminals on 4 channel models (±12 V each) does not exceed 2.0 A.
Otherwise, the instrument or a device connected to the probe power terminals may break.
French
ATTENTION
Ne pas utiliser les bornes d'alimentation de la sonde (en option) sur le panneau latéral droit
de l'instrument à des fins autres que l'alimentation d'une sonde FET, d'une sonde de courant
ou d'une sonde différentielle. S'assurer que le courant total des huit bornes d'alimentation
de la sonde et les huit bornes d'interface de la sonde (± 12 V chacune) sur les modèles à 8
canaux ou celui des quatre bornes d'alimentation de la sonde et les quatre bornes d'interface
de la sonde sur les modèles à 4 canaux (± 12 V chacun) ne dépasse pas 2,0 A. Sinon,
l’instrument ou un appareil connecté aux bornes d’alimentation de la sonde pourraient casser.
2.4 Connecting Probes