App-13
IM DLM3054-03EN
Appendix
1
2
3
4
5
6
App
Index
When the Set Record Length Is 500 Mpoint
(Selectable on models with the /M2 option)
Mode
Setting
High Resolution Mode OFF
High Resolution Mode ON
RealTime
Interpolation
Repetitive
RealTime
Interpolation
Repetitive
Time/div
Sample
Rate
(S/s)
Display
Record
Length
(Point)
Sample
Rate
(S/s)
Display
Record
Length
(Point)
Sample
Rate
(S/s)
Display
Record
Length
(Point)
Sample
Rate
(S/s)
Display
Record
Length
(Point)
Sample
Rate
(S/s)
Display
Record
Length
(Point)
Sample
Rate
(S/s)
Display
Record
Length
(Point)
500s
100k
500M
Interpolation mode and Repetitive
sampling mode cannot be used with
this record length.
125k
500M
Interpolation mode and Repetitive
sampling mode cannot be used with
this record length.
200s
250k
500M
250k
500M
100s
500k
500M
500k
500M
50s
1M
500M
1M
500M
20s
2.5M
500M
2.5M
500M
10s
5M
500M
5M
500M
5s
10M
500M
10M
500M
2s
25M
500M
25M
500M
1s
50M
500M
50M
500M
500ms 100M
500M
100M
500M
200ms 250M
500M
250M
500M
100ms 500M
500M
500M
500M
50ms 625M
312.5M
625M
312.5M
20ms 2.5G
500M
1.25G
250M
10ms 2.5G
250M
1.25G
125M
5ms
2.5G
125M
1.25G
50M
2ms
2.5G
50M
1.25G
25M
1ms
2.5G
25M
1.25G
12.5M
500µs 2.5G
12.5M
1.25G
6.25M
200µs 2.5G
5M
1.25G
2.5M
100µs 2.5G
2.5M
1.25G
1.25M
50µs
2.5G
1.25M
1.25G
625k
20µs
2.5G
500k
1.25G
250k
10µs
2.5G
250k
1.25G
125k
5µs
2.5G
125k
1.25G
62.5k
2µs
2.5G
50k
1.25G
25k
1µs
2.5G
25k
1.25G
12.5k
500ns 2.5G
12.5k
1.25G
6.25k
200ns 2.5G
5k
1.25G
2.5k
100ns 2.5G
2.5k
1.25G
1.25k
50ns
2.5G
1.25k
1.25G
625
20ns
2.5G
500
1.25G
250
10ns
2.5G
250
1.25G
125
5ns
2.5G
125
1.25G
62.5
2ns
2.5G
50
1.25G
25
1ns
2.5G
25
1.25G
12.5
Acquisition when you specify this record length is performed in single mode.
: Roll mode