Item
Specifications
Immunity
Compliant standards
EN 61326-1 Table 2 (for industrial environments)
(Applies to the DLM3022, DLM3024, DLM3032, DLM3034, DLM3052, DLM3054)
EN 61326-2-1
Influence in the immunity testing environment (criteria A)
Noise increase:
500 μV/div to 10 mV/div: ±4 div
20 mV/div to 10 V/div: ±2 div
Test conditions:
2.5 GS/s, envelope mode
Cable conditions:
Same as the emission cable conditions
Test items:
1. Static discharge: EN 61000-4-2
Air discharge: ±8 kV. Contact discharge: ±4 kV. Criteria B.
2. Radiated immunity: EN 61000-4-3
80 M to 1 GHz, 10 V/m, 1.0 GHz to 6.0 GHz, 3 V/m, Criteria A
3. Conducted immunity: EN 61000-4-6
3V, Criteria A
4. EFT/Burst: EN 61000-4-4
Power line: ±2 kV. Signal line: ±1 kV, Criteria B
5. Surge immunity: EN 61000-4-5
±1 kV between lines, ±2 kV common, Criteria B
6. Voltage dip and interruptions: EN 61000-4-11
1 cycle, both polarities, 100%, criteria B
Other tests, criteria C
Definitions of criteria A, B, and C
Criteria A: During testing, "influence in the immunity testing environment"
described above is met.
Criteria B: The instrument continues to function and is controllable
throughout testing. The instrument does not change operation
modes, and data changes do not persist.
Criteria C: During testing, temporary degradation of performance or loss
of functionality occurred, the correction of which required user
operation or system reset.
Environmental standard
Compliant standards
EN 50581 Monitoring and control instruments including those for industrial use
6.11 General Specifications