31
■
Selecting the Wavelength Using Detailed Values (in 5 nm Steps)
Step
Key
Display
Description
1 SETUP
is displayed.
The light receiving mode indicator
blinks.
Changes to light
receiving mode
settings.
2 SETUP
The wavelength setting blinks (ex.
“850” blinks), and the light
receiving mode stops blinking.
Changes to
wavelength
settings.
3 [UP]
or
[DOWN]
850nm => 1300nm => 1310nm =>
1490nm =>1550nm => 1625nm
=> 1650nm => previous value
=>
USR
=> 850nm
Press the key
repeatedly to scroll
through the options
([DOWN] scrolls in
reverse).
Select USR.
4 ENTER
is displayed to the lower
right of . The wavelength
setting blinks (ex.: “1315”)
Changes to the
wavelength user
settings.
5 [UP]
or
[DOWN]
750nm => 755nm => 760nm
=>765nm =>
・・・・
1690nm =>
1695nm =>1700nm
Changes in 5 nm
steps.
Hold the key down
to scroll rapidly.
6 ENTER The mode display stops blinking.
ON or OFF blinks.
Selects the
measurement
wavelength.
Changes to the
averaging setting.
7 ENTER
changes to the measured
value display.
Exits measurement
wavelength
settings.
<Display Example>
Wavelength Setting USR Screen (Step 3) 5 nm Step Wavelength
Display Screen (Step 6)