21
2-5-5 System Reset
This function restores settings such as measurement conditions to their
original condition upon shipment from the factory.
Procedure
Display
Description
Turn the power
ON while
pressing SETUP.
CLr and rESU are displayed,
followed by measured
values.
Start-up procedure.
Hold down until screen
displayed.
This function restores the Resume values, and does not initialize the
standard and user memories. After resetting the system, the
instrument enters normal measurement mode in which the system
reset values are used as the initial values.
<Display Example>
System Reset Screen
<Initial Values upon System Reset>
Section
Initial Value
Mode, changing to normal
measurement
Absolute value meas.
Measurement wavelength
1310 nm
Measurement mode
CW
Displayed units
dBm
Averaging OFF
User calibration function setting
OFF
User calibration value
0.00 dB
Standard memory address
000
User memory address
U00
Sort key
0
Even when starting with a system reset, Power save is set if
powered by batteries. However, if you hold down the POWER
key after executing a system reset, the Power save setting is
disabled.