12
SETUP
Setup Mode Key
Press this key to change to the Setup mode, in which various settings
can be entered. In absolute value measurement mode, measurement
conditions can be entered as follows.
• Light receiving mode setting (switching of the modulation frequency for
CW light and chopped light)
• Wavelength settings (typical value, or switching of the wavelength in 5
nm steps)
• Averaging settings
Make selections using the UP/DOWN key, then press ENTER to enter
the selection. If no changes are made, press the ENTER key
repeatedly until you return to absolute value measurement mode. In
relative value measurement mode you can set an arbitrary reference
value for relative value measurement (dBm). Select the reference
value using the UP/DOWN key, then press ENTER to enter the
selection. After entering a setting, the instrument returns to absolute
value measurement mode. (Related keys: UP/DOWN, ENTER)
ENTER Entry
Key
Enters various kinds of selectable settings such as addresses. They
key also clears various types of errors (data saving, data deletion, and
resume errors).
Special operation: Hold down this key while pressing
another key
. You can hold down the ENTER key while pressing the
MEMORY key to perform user memory measurement. Also, if you hold
down the ENTER key while pressing the SETUP key, you can execute
the user calibration function.
(Related keys: UP/DOWN, SETUP, MEMORY)
▲
,
▼
Selection Key (UP/DOWN)
Selects various kinds of settings such as addresses. Press once to
change the selection one item UP or DOWN. Hold down to quickly
scroll UP or DOWN through the selections. (Related keys: SETUP,
MEMORY, ENTER)
MEMORY
Memory Measurement Key
Pressing this key places the instrument in memory measurement mode.
In memory measurement mode, operations related to memory data can
be performed in the following order.
• Memory measurement (save measured data to an arbitrary address as it
is being measured).
• Memory recall (view any saved data in detail, recall reference values or
measurement conditions)
• Memory delete (delete any saved data or all data)
If no operation is to be performed, you can press the MEMORY key
repeatedly until the instrument returns to normal measurement mode.