24
3-1-2 Attaching a Connector Adapter
A specialized connector adapter (SU2004A-***, optional) is required
for connecting an optical connector to the instrument. Choose an
adapter that matches the optical connector you will use.
The screw mechanism that ataches the connector adapter to the
instrument is of precision manufacture. Please handle all parts
with care so as not to damage the screw threads.
Do not allow dirt to come into contact with the optical input
section. Dirt or other foreign particles can influence
measurement. Attach the protective cap to the optical input
connector when not in use.
(1) Remove the optical input connector cap.
(2) Attach a connector adapter (SU2004A-***, optional).
The connector adapter screws onto the connector; hold the
adapter upright to align the threads properly, then turn to
tighten.
Do not attach the connector adapter at an angle, and do not
forcibly turn it any further after it has initially tightened. Doing so
can damage the screw threads or sensor.
It is recommended to clean the inside of the connector adapter
using a dedicated swab or other cleaner.
Dedicated swabs for cleaning optical adapters are available from
NTT, including the “OPTIPOP-S” (NTT-AT), and the “CLETOP
Stick-Type” (NTT-ME).