3-3 Entering Measurement Conditions
The following three measurement condition parameters can be set.
1) Light receiving mode setting (switching of the modulation
frequency for CW light and chopped light)
2) Wavelength settings (typical value, or setting of the wavelength
in 5 nm steps)
3) Averaging settings
These three measurement conditions can be set all at once, but
the following explains how to set each one individually.
The conditions set here are stored until the the next time you enter
settings, even if the power is turned OFF (Resume Function).
The setting procedures described all start from the Initial state
(absolute value measurement in normal measurement mode).
<Display Example> Initial state
Same screen as when first starting up.
Confirm that the screen shows the
Initial state, then begin the procedure
for entering measurement conditions.
PWR SAVE
nm
CW
dBm
PWR SAVE
PWR SAVE
nm
CW
dBm
If you make a mistake, press the ESC or SETUP key.
ESC key: Return to normal measurement mode.
SETUP key: Cancel the setting and advance to next item.