27
3-2 Turning the Power Supply ON and OFF
3-2-1 Turning ON the Power
■
Starting Up in Power Save Mode
When the power is OFF, press POWER/ESC.
is displayed in the upper left part of the screen.
<Display Example>
In Power save mode, the power turns
OFF if no key is pressed for ten
minutes.
However, Power save mode is
disabled when using the AC adapter.
■
Starting Up When Power Save Mode Is OFF
When the power is OFF, hold down POWER/ESC until
disappears from the screen.
The AC adapter is available as an option (SU2007A-*).
Never use an AC adapter or power cable other than the
dedicated one available for use with this instrument, as damage
can result.
* This instrument performs a zero adjustment (auto-offset) when
the power is turned ON.
For low level measurement (-50 dBm or less), in order to cancel
out temperature changes and other effects, it is recommended to
power cycle the instrument before measurement to perform the
zero adjustment. The zero adjustment need not be performed
while shielded from light. The optical connector cap is designed
to protect the connector from dirt and collision, and does not
provide light shielding.
3-2-2 Turning the Power OFF
Hold down the power key until the LCD display goes out.
Remove the optical connector and attach the optical connector
cap.
When using the AC adapter, turn the instrument OFF with the
power key before connecting the adapter.