10
2. PRE-OPERATION PROCEDURES
Installation
3. Make water supply connection.
(1) Remove rubber cap from connector
plug
⑧
.
(2) Insert connector plug
⑧
securely
into connector socket
⑨
with the
sleeve held down in direction of
arrow. Sleeve returns to the
original position when released, If
connection is secure.
※
The connector socket contains a
valve, which opens only when
socket is inserted. When socket is
removed, water cannot enter.
4. Connect water supply hose only to tap equipped with sink, where possible.
If water supply hose is connected to a
water tap without sink facilities,
flooding may occur if water supply
hose is unintentionally disconnected or
damaged. Connecting to a water tap
having a sink is therefore preferable.
5.
If no sink is available.
●
Use optional Water Supply Port Unit (found on P.30) designed to provide a more secure
connection under fluctuating water pressure than the standard connector kit and hose.
6. If no tap is available.
●
Use one of the appropriate optional Water Supply Port Coupler Joints on P.30.
Sleeve
⑧
Plug
Valve
⑨
Socket
Tap
Water supply
hose
Connector
Connection port Body
Sink
Summary of Contents for Pure Line WE200