4. OPERATION PROCEDURE
Water Quality Index
Changing water quality index
The water quality index may be changed by
pressing UNIT .
●
Conductivity reading
●
Resistivity reading
※
Index conversion formula is shown below:
Conductivity
(S/m)
=
1
Resistivity (
Ω
・
m)
=
1
Resistivity (
Ω
・
m)
Conductivity (S/m)
See
“How conductivity/resistivity relates to water purity” (P.27) for details.
Reverse Osmosis (RO) Membrane Flush
If ELB is ON, the RO membrane flush function automatically begins six hours after last
sample draw or 24 hours after last system flush.
●
Control panel displays will appear as shown
on the left, during RO membrane flush.
●
Flush duration is about one minute.
※
Samples may be taken during RO membrane
flush and resumes automatically when
sample is finished dispensing.