5. OPTIONAL ITEMS & MODIFICATIONS
Optional Remote Sampling Terminal
The remote sampling terminal allows sampling with an input device from a remote location.
Connection terminals
(Terminal block screw size
:
M4)
Example: Connect and use a switch.
Connect a switch or other input device to the
remote sampling input terminals, located on right
panel of main unit.
Terminal block:
REMOTE
Input contact: Non-voltage
input terminal
・
Remote sampling procedure:
Sample is dispensed when contacts (circuit) are
closed by switch or other device.
・
Sample stops being dispensed when contacts are
opened.
※
Do not apply voltage to the remote sampling
input terminal. Unit malfunction may result.
PUSH key is disabled during remote sampling.
※
MIZU guard (product code
:
254199) is
recommended for added safety in remote
installations.
Optional Remote Alert Terminal
The external alarm function generates an alarm signal when unit errors occur or when wear items
need replacement.
Connection terminals
(Terminal block screw size
:
M4)
Connect pilot light or other alert devices to external
alarm output terminals, located on right panel of
main unit.
Terminal block
:
ALARM Output circuit
:
Relay
a-contact (Normal
Open)
Contact capacity
:
AC250V 1A
(
Resistance load
)
DC30V
1A
(
Resistance load
)
Connection load example:
Connect alarm lamp or alert buzzer..
・
Relay contact closes when alarm signal is
generated.
・
External alarm signal will continue to be
generated, even if alarm is cancelled.
・
All external alarm signals, excluding those for
leaks, stop once power key is turned OFF.
REMOTE
(input)
Connection
load
ALARM
REMOTE
ALARM
(output)
Load power
supply