6. MAINTENANCE PROCEDURES
Ion Exchange Resin Cartridge Replacement
For replacement procedure, see
“Install the ion exchange resin cartridge” (P.11). Notification alarm will
automatically reset once water quality improves, following cartridge replacement.
・
Degraded water quality and processing capacity will be exhibited from cartridges stored for an extended
time period. Discretion is therefore advised in preparing spare cartridges. Cartridge shelf life is
approximately 4 months.
・
Spent cartridges should be discarded as non-combustible material or returned to Yamato (applicable in
Japan only). Use the mailing label supplied with cartridge for returns. Yamato Scientific promotes
collection and recycling for preservation of the environment.
・
Following cartridge replacement, discard the first 4
ℓ
of water dispensed to purge system of air and initial
impurities.
Pre-process Cartridge Replacement
For replacement procedure, see
“8. Install the pre-process cartridge” (P.12). To reset notifications/alerts, see
“How to reset after replacement of consumable parts” (P.36).
・
Ion exchange resin cartridge life will shorten if unit is used without replacing the pre-process cartridge at
specified interval.
・
Spent cartridges should be discarded as non-combustible material or returned to Yamato (applicable in
Japan only). Use the mailing label supplied with cartridge for returns. Yamato Scientific promotes
collection and recycling for preservation of the environment.
Reverse Osmosis (RO) Membrane Replacement
“13. Install the reverse osmosis (RO) membrane.
To reset notifications alerts, see
“Wear Item Replacement Alert Reset” (P.36). Running Initial wash and
adjusting the pressure calibration valve must be done after replacing RO membrane. For initial wash pressure
calibration procedures, see
“Run initial wash for reverse osmosis membrane” (P.16) and “15. Calibrate
pressure for reverse osmosis membrane.
(P.17).
・
Ion exchange resin cartridge life is shortened and specified water quality and/or sample size will not be
accurate if RO membrane is not replaced at specified interval.
・
Spent cartridges should be discarded as non-combustible material.