37
4. Operating procedures
Useful functions (RS485 communication function)
7. Identifier
/
command list
<
About identifiers and settings
>
*1
:
When the time exceeds 100 hours, setting unit will be one hour.
*2
:_
indicates a space.
*3
:
Parameters for which W command is effective during each operation mode. (This is
effective during start phase in the normal mode.)
Fixed Temp. operation parameters
Name
Identifier
Command
Setting
Temp. setting
SV1
R/W/*/*
SLL
~
SLH
:
Lower limiter
~
upper limiter
℃
*3
Event output setting
EVE
R/W/*/*
00000
:
Event output OFF
00001
~
00015
:
1
~
15 patterns
Motor output setting
FAN
R/W/*/*
For motor output operation selection,
When ON/OFF control is selected
00000
:
Motor control disabled
00001
:
Motor control enabled
When level control is selected
00000
:
Motor control disabled
00001
~
00010
:
Levels 1
~
10
Refrigerator output setting
LEI
R/W/*/*
00000
:
Refrigerator control disabled
00001
:
Refrigerator control enabled
(
normal operation
)
00002
:
Refrigerator control enabled
(
forced operation
)
Program operation parameters
Name
Identifier
Command
Setting
Temp. setting
S01
~
S99
R/W/*/*
SLL
~
SLH
:
Upper limiter
~
lower limiter
℃
*1*2
Time setting
T01
~
T99
R/W/*/*
00000
~
09959
:
0hr 0min
~
99 hrs 59min
H0000
~
H9999
:
100hrs
~
9999hrs
Repeat dest in. setting
R01
~
R99 R/W/*/*
00001
~
00099
:
Steps 1
~
99
Repeat number setting
C01
~
C99 R/W/*/*
00000
:
None
00001
~
00099
:
1
~
99 times
00100
:
No restriction
End operation setting
O01
~
O99 R/W/*/*
00000
:
END status
00001
:
HLD status
00100
:
FT status
Pattern
№
setting
PSN
R/W/*/*
00001
~
00099
:
Programs 1
~
99
No of step setting
STC
R/W/*/*
00000
:
No steps
(
unregistered
)
00001
~
00099
:
Steps 1
~
99
Program
№
select
PSP
R/W/*/*
00000
:
Program 1
00001
:
Program 2
00100
:
Program 3