27
4. Operating procedures
Useful functions (Temperature output terminal)
Specifications
Temperature output
(
ANALOG
)
Outputs a voltage (DC) according to the measured temperature
Output temperature range: IL702 -5
~
65
℃
Output voltage: DC1
~
5V
Resolution
: 0.1
℃
Connection
: M4 screw terminal block
Alarm output
(
ALARM
)
Outputs when an abnormality is detected (For description of
abnormalities, see
“Safety device and error codes” on P. 47.)
A-contact (relay contact)
Contact capacity AC250V 3A (resistance load)
DC30V
3A (resistance load)
Connection
:
M4 screw terminal block
Time up output
(
TIME UP
)
Outputs on time-up of AUTO STOP, AUTO START or QUICK
AUTO STOP or at program end.
A-contact
(
relay contact
)
Contact capacity: AC250V
3A
(
resistance load
)
DC30V
3A
(
resistance load
)
Connection
:
M4 screw terminal block
IL702
Temperature
(℃)
Output voltage
(
V
)
-5
1.00
5
1.57
15
2.14
25
2.71
35
3.29
45
3.86
55
4.43
65
5.00