4. Operating procedures
Operating sequence (Basic operation)
5. Set the number of steps.
Then enter the number of steps. Enter a number of steps you want. When finished, press the
ENTER key.
※
Take steps for heating or cooling into consideration beforehand. Set
“TEMP.” you want to
“1 min”. The setting “1 min” for the timer parameter means you can move to the next step
only after the unit has reached the temperature you want at the maximum power. So set
the time parameter for heating or cooling steps to
“1 min”.
6. Set a step temperature.
The SET TEMP. screen flashes and the mode switches to the temperature setting mode.
Select a temperature you want with the
▽△
keys to change to it. When finished, press
the ENTER key.
7. Set the step time.
The TIMER screen flashes and the mode switches to the time setting mode. Select a time
you want with the
▽ △
keys to change to it. When finished entering, press
the ENTER key.
※
For heating/cooling steps
→
”1 min"
For holding step
→
”Time you want to hold”
8. Set each of steps.
Repeat steps 7.
~
8. to set each step to the temperature and time you want.
9. END setting
The CR5 controller has a function that allows selecting end operations from below. Make
selection below during the program setting phase. The number of digits in the SET TEMP.
screen changes to four and the lowest digit flashes.
Select any of
“0000
:
OFF
”, “0001
:
Hold
”, “0002
:
Fixed temp.
” with the
▽△
keys and press
the ENTER key. Now program setting has completed.
0
:
OFF
Terminates program operation and brings the unit into the standby state.
1
:
HLD
An alarm notifies the end of program operation, display END, and control
operation is completed at once.Also, the elapsed counts is displayed since
control end.
2
:
ST
After completion of the final step, the unit automatically switches to the Fixed
Temp. operation mode and continues operation.
Make changes following the program setting steps above.