25
4. Operating procedures
Operating sequence (Example of program registration)
1. Register a name
“37
℃
24
4
℃
72
”.
Perform steps 1.
~
4. on P.21
”, section 3-3
)
, then perform step 5.
and enter
“37
℃
24
4
℃
72
”.
PROGRAM01 EXIST
NAME (37
℃
24
4
℃
72)
2. Enter the number of steps.
Set "4" as the number of steps.
(Settings for step 1
~
step 4.)
STEP in PGM01
SET:** REMAIN:**
3. Step 1
:
Heating to 37
℃
and then set the temperature parameter.
Change the temperature flashing in the SET TEMP. screen to
37.0
℃
with the
keys. Press the ENTER key to determine
the temperature.
PROGRAM01 Step 01
TEMP SET
4. Step 1
:
Heating to 37
℃
and then set the time parameter.
Change the time flashing in the TIMER screen to 00:01 with the
keys. Press the ENTER key to determine the
temperature.
PROGRAM01 Step 01
TIME SET
5. Step 2
:
Hold at 37
℃
for 24 hrs and then set the temperature
parameter.
Change the temperature flashing in the SET TEMP. screen to
37.0
℃
with the
keys. Press the ENTER key to determine
the temperature.
PROGRAM01 Step 02
TEMP SET
6. Step 2
:
Hold at 37
℃
for 24 hrs, and then set the time time
parameter.
Change the time flashing in the TIMER screen to 24
:
00 with the
keys. Press the ENTER key to determine the
temperature.
PROGRAM01 Step 02
TIME SET
7. Step 3
:
Heating to 4
℃
and then set the temperature parameter.
Change the temperature flashing in the SET TEMP screen to
4.0
℃
with the
keys. Press the ENTER key to determine
the temperature.
PROGRAM01 Step 03
TEMP SET
8. Step 3
:
Heating to 4
℃
and then set the time parameter.
Change the time flashing in the TIMER screen to 00
:
01 with the
keys. Press the ENTER key to determine the
temperature.
PROGRAM
01 Step 03
TIME SET
9. Step 4
:
Hold at 4
℃
for 72 hrs and then set the temperature
parameter.
Change the temperature flashing in the SET TEMP. screen to
4.0
℃
with the
▽△
keys. Press the ENTER key to determine
the temperature.
PROGRAM 01 Step 04
TEMP SET
10. Step 4
:
Hold at 4
℃
for 72 hrs and then set the time parameter.
Change the time flashing in the TIMER screen to 72
:
00 with the
keys. Then press the ENTER key to determine the
temperature.
PROGRAM 01 Step 04
TIME SET
11. Set the program end operation.
In the previous section, pressing the ENTER key will change
the screen to the Program End operation screen. Set the value in
the SET TEMP. screen to
“0001” and press the ENTER key.
Now setting has been completed.
PGM01 END SET
0
:
OFF
1
:
HLD
2
:
FT
PROGRAM 01 EXIST
Name
(
37
℃
24
4
℃
72
)