YCO-201008
DP-6700
Technical Manual
1.3.6 Span adjustment (calibration)
Related pages:
Page 1-5 1.3.1 How to enter test mode and meanings of panel indications
Make sure to use weights that have the accuracy of Class 2 standard weight or highter, and carry out
the following procedure for span adjustment.
Get the weights for 500 increments, 1/2 weighing capacity and full weighing capacity to perform span
adjustment. For example, to calibrate the scale with weighing capacity of 30kg, 2 sets of weights total-
ing 15kg and a 5000g weight are necessary.
Caution: Span adjustment may not be made properly in a place exposed to vibrations or wind therefore
avoid such places.
1-24
Chapter 1 Adjustment method
Summary of Contents for DP-6700
Page 2: ...YCO 201008 DP 6700 Technical Manual ...
Page 28: ...YCO 201008 DP 6700 Technical Manual 1 25 Chapter 1 Adjustment method ...
Page 29: ...YCO 201008 DP 6700 Technical Manual 1 26 Chapter 1 Adjustment method ...
Page 30: ...YCO 201008 DP 6700 Technical Manual 1 4 Wiring diagram 1 27 Chapter 1 Adjustment method ...
Page 32: ...YCO 201008 DP 6700 Technical Manual 1 29 Chapter 1 Adjustment method ...
Page 33: ...YCO 201008 DP 6700 Technical Manual 1 30 Chapter 1 Adjustment method ...
Page 38: ...YCO 201008 DP 6700 Technical Manual Chapter 3 How to disassemble 3 1 ...
Page 39: ...YCO 201008 DP 6700 Technical Manual Chapter 3 How to disassemble 3 2 ...
Page 40: ...YCO 201008 DP 6700 Technical Manual Chapter 3 How to disassemble 3 3 ...
Page 41: ...YCO 201008 DP 6700 Technical Manual Chapter 3 How to disassemble 3 4 ...