YCO-201008
DP-6700
Technical Manual
1-17
Chapter 1 Adjustment method
B0(110)
Factory setting
000:
001:
002:
003:
004:
005:
006:
007:
Not use default setting
DP-6700 30kg/10g, 1/3000, Verified Model
DP-6700 60kg/20g, 1/3000, Verified Model
DP-6700 150kg/50g 1/3000, Verified Model
DP-6700 150g/100g, 1/1500, Verified Model
DP-6700 30kg/10g, 1/3000, Non-Verified Model
DP-6700 60kg/20g, 1/3000, Non-Verified Model
DP-6700 150kg/50g, 1/3000, Non-Verified Model
B5(115)
Transmission of ad value
000:
001:
002:
003:
004:
Not send
Raw (2 batches average) + 4th filtering
Raw (2 batches average) + Final filtering
4th fil Final filtering
4th fil Final filtering at very stable - 20
B6(116)
Input judgment of filter "L" at
stable
000-255
By 0.01% for capacity
B7(117)
Input convergence of filter "S"
at stable
000-015:
By 0.01% for capacity
B8(118)
Input difference of filter "S" at
stable
000-015:
By 0.01% for capacity
B9(119)
Auto tare delay, subtracting
grading
000-099:
Delay time (x10ms)
C7(127)
Keyboard tare
000:
001:
Keyboard tare is invalid
Keyboard tare is valid
C8(128)
Recommended calibration
method
002:
003:
004:
005:
006:
007:
Using 2 points - zero point and full capacity
Using 3 points - zero, 1/2 cap, and full capacity
Using 4 points - zero, 1/2 cap, full cap and 1/2 cap on return
Using 4 points - zero, 500 increments, 1/2 cap and full cap
Using 4 points - zero, 500 increments, 2/3 cap and full cap
Using 4 points - zero, 1/3 cap, 2/3 cap and full cap
C9(129)
Store display hold mode
000-001:
Memorize display mode and revert to current status at power
on next time
How to Set Factory Parameters
Description
Indication
Display the internal count in test mode referring to 1.3.1 How
to enter test mode and meanings of panel indications on
Page 1-5. When the internal count is shown on the panel,
press the
key while holding down the
key; and it
switches to span adjustment mode. (See 1.3.6 Span adjust-
ment (calibration) on Page 1-22)
In span adjustment mode, press
key while holding down
the
key; and it switches to system parameter mode.
Going to the next parameter:
Press the
key to go to the next parameter.
To change the setup value, enter a new value and go to the
next parameter. Make sure to go to the next parameter to
enable the new value. If you exit before moving to the next
parameter the system discards the enter valve
Summary of Contents for DP-6700
Page 2: ...YCO 201008 DP 6700 Technical Manual ...
Page 28: ...YCO 201008 DP 6700 Technical Manual 1 25 Chapter 1 Adjustment method ...
Page 29: ...YCO 201008 DP 6700 Technical Manual 1 26 Chapter 1 Adjustment method ...
Page 30: ...YCO 201008 DP 6700 Technical Manual 1 4 Wiring diagram 1 27 Chapter 1 Adjustment method ...
Page 32: ...YCO 201008 DP 6700 Technical Manual 1 29 Chapter 1 Adjustment method ...
Page 33: ...YCO 201008 DP 6700 Technical Manual 1 30 Chapter 1 Adjustment method ...
Page 38: ...YCO 201008 DP 6700 Technical Manual Chapter 3 How to disassemble 3 1 ...
Page 39: ...YCO 201008 DP 6700 Technical Manual Chapter 3 How to disassemble 3 2 ...
Page 40: ...YCO 201008 DP 6700 Technical Manual Chapter 3 How to disassemble 3 3 ...
Page 41: ...YCO 201008 DP 6700 Technical Manual Chapter 3 How to disassemble 3 4 ...