XO 4 & XO FLEX TROUBLE SHOOTING GUIDE
Version 3.30
51
If 2 USB devices are connected,
they will get 250 mA each which
is NOT sufficient for the Vistacam
IX.
Check the camera is getting 500
mA power supply.
If the camera is connected to a
USB group on the dental
computer, the camera might be
sharing the power supply which
is insufficient.
Connect the camera to a USB
hub with external power supply.
Requirements are:
minimum 480mbit/s
minimum 500 mA
Missing image/connection
Bypass the unit and cables in the
floor by, and check camera is
working by connecting the
camera directly to a computer.
Use test cable CA-220 or AP-071
+ CA-084.
Check the camera and
cables/connection by connecting
a different computer to the unit.
Bypass cables in the floor and
check the camera is working by
connecting a computer directly
to the USB cable in the bridge
arm.
Missing image/connection
due to too long cables
Use an active USB cable.
CA-120 XO cable kit.
The camera is active when
the suspension is at rest.
The magnet not aligned or not
correctly placed in the field of the
hall contact.
Align the suspension.
Check that the magnet is present
and properly aligned. If the
magnet is missing, change the
AN-011.
If not aligned, check that the
screws on the suspension
bearing bracket are secure.
Suspension cable is not mounted
properly or is defective.
Check that the CA-101 cable
connecting the AN-368 PCB to
the AO-329 for vistacam iX or the
AO-347 for iX HD PCB is
properly connected.
No Light
Camera is defective
Replace the camera
Suspension PCB is defective
Change Suspension PCB
AO-329 : Vistacam iX
AO-347 : Vistacam iX HD
Summary of Contents for 4
Page 1: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE YB 907...
Page 30: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 30 TECHNICIAN MENU Configuration Adjustments...
Page 31: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 31 SUSPENSION CONFIGURATION...
Page 54: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 54...
Page 58: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 58 LIQUID SENSOR PCB AP 869...
Page 64: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 64 SUCTION DISINFECTION SYSTEM...