XO 4 & XO FLEX TROUBLE SHOOTING GUIDE
Version 3.30
10
START UP
Failure description
Reason
Solution
No “welcome” message on
display.
No light on the main switch
(early model)
Main switch OFF
Activate the main switch
Main switch ON
Check LED D1 and D3 on Mains
PCB are “ON”
Check the mains supply voltage
at J4 on Mains PCB, 230VAC.
Replace fuse(s) on the Mains
PCB if necessary.
Check the mains switch.
No light in the display
If the unit is ON and the Mains
fuses are ok
After removing the instrument
bridge top cover, check the
following
Check that diodes D800, D191
and D110 at Bridge PCB are ON.
If LED D110 (3.3VDC) is ON
Check CA-103 Display cable for
visible defects
Check that CA-103 at J12
connector is properly connected.
Check for corrosion at Display
PCB from leaking water
Measure voltage on J12:
Between pin 1 and 5 = 5 V
Between pin 9 and 10 = 3.3 V
Replace the Display PCB
If LED D110 is OFF.
Replace Bridge PCB.
No “Welcome” message on the
display and no sound during
start up.
Check LED D1, D2, D4 and D6
on the Power supply PCB
(replace fuses if necessary)
Verify all connections in CA-004
– Bridge power cable
For more information, click on
the following link “Bridge PCB”
on page 19, “Power Supply
PCB” page 14 and “Mains PCB”
page 13.
Summary of Contents for 4
Page 1: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE YB 907...
Page 30: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 30 TECHNICIAN MENU Configuration Adjustments...
Page 31: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 31 SUSPENSION CONFIGURATION...
Page 54: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 54...
Page 58: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 58 LIQUID SENSOR PCB AP 869...
Page 64: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 64 SUCTION DISINFECTION SYSTEM...