XO 4 & XO FLEX TROUBLE SHOOTING GUIDE
Version 3.30
28
“SUSPENSION PCB”, AD-605.
Different board designs based on AO-605:
The board function provides the voltages required to operate the instruments (voltage, water and air),
supplies the voltage to the optical fibre and via the Hall effect sensor to detect if the instrument is
activated.
Failure description
Reason
Solution
LED D110 remains OFF
when the instrument is
pulled forward.
Suspension cable is not properly
mounted or is defect.
Check suspension cable is firmly
connected
Replace suspension cable.
Suspension PCB defect
Replace Suspension PCB.
Wrong instrument.
Displayed if the instrument in
hand does not match the
suspension configuration or is
placed on the wrong connection
Before entering the technician
menu.
Verify that the instrument is
properly connected.
Verify that the instrument is
inserted on the correct
connector.
Verify the number behind the
connector.
In the technician menu, Verify the
configuration in the “SUSP 1-6”
and set the correct item.
Consult the “Technician menu”
on page 26 for more information.
If the parameters described
above are correct:
Measure the 3.3V on pin 15 J2x
(J22 to J26) when the instrument
is activated. If the 3.3V is not
present, the main board "AN-
368" is defective, change it.
If the 3.3V is present it indicates
that the Suspension PCB is
defective, change it.
Wrong Susp. PCB.
Displayed if the instrument being
handled is not compatible with
the suspension PCB.
Before configuring the SPCB,
check the table “Suspension
Configuration” under “to be
selected in SPCB 1-6” to find the
correct PCB.
Configure “SPCB 1-6” in the
technician menu.
Summary of Contents for 4
Page 1: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE YB 907...
Page 30: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 30 TECHNICIAN MENU Configuration Adjustments...
Page 31: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 31 SUSPENSION CONFIGURATION...
Page 54: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 54...
Page 58: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 58 LIQUID SENSOR PCB AP 869...
Page 64: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 64 SUCTION DISINFECTION SYSTEM...