Information:
“
WRONG SUSP PCB
” also occurs after an update of the unit's firmware or after a “
Reset All
”
has been performed and is especially relevant for MX2 micromotors.
The MX2 micromotor does not work if the parameter “
AD-605
” in the “
SPCB 1-6
” menu is
selected; “
WRONG SUSP PCB
” will be displayed when the instrument is activated.
After an update of the Unit or a “
Reset All
”, the “
AD-654
” settings are reset to “
AD-605
”.
To reconfigure the parameter, enter the “
SPCB 1-6
” menu, select the correct position where
the micromotor is located and select “
AD-654
” from the menu.
The turbine and/or the MC3 type micromotor operates correctly when the parameter
“
AD-
605
”
in the “
SPCB 1-6
” menu is selected.
However, if the MC3 micromotor or turbine is equipped with an LED and the parameter is set
to “
AD-605
”, this may have an impact on the life of the LED.
Reconfigure the parameters in the
“
SPCB 1-6
”
menu and select
“
AD-654
”
Setting AD-654 will set max Voltage to 3.0V
MC3 micromotors and/or turbines equipped with halogen lamps. Works perfectly under the
parameter
“
AD-605
”
in the menu
“
SPCB 1-6
”
. It is also possible to set the parameter to
“
AD-
654
”
NB: The LED light intensity on the MX2 motor cannot be modified Always 3.0V
It is possible to vary the halogen light intensity in the menu “
FIBER
” by setting the parameter
from 0V to 3.3V when configured “
AD-605
” and 0V to 3.0V when configured “
AD-654
”
Voltage measurement on the suspension PCB connector (CA-100, CA-101...)
Summary of Contents for 4
Page 1: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE YB 907...
Page 30: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 30 TECHNICIAN MENU Configuration Adjustments...
Page 31: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 31 SUSPENSION CONFIGURATION...
Page 54: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 54...
Page 58: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 58 LIQUID SENSOR PCB AP 869...
Page 64: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 64 SUCTION DISINFECTION SYSTEM...