XO 4 & XO FLEX TROUBLE SHOOTING GUIDE
Version 3.30
26
“PROPORTIONAL VALVE”, AN-903:
Purpose:
The proportional valve controls the amount of water distributed to all instruments except the syringe.
The proportional valve is fully controlled by PWM pulse on a constant frequency of 78 kHz by the
“Power driver” AO-137 PCB on J5.
Spare parts for AN-903
3 Gasket
MR-150
4 O-ring,12 x 1,5 B Viton
SD-307
Green tube dental.
FS-078
Failure description
Reason
Solution
No water on instruments.
No water. Fault on waterline
(instrument bridge)
Check if the water reaches the
syringe 3F/6F
The flow meter nozzle is
obstructed
Replace the flow meters nozzle
(MG-608)
For more information, go to Page
23 “Flow meter”.
The “Power driver” is defect.
Change “Power driver” AO-137
on page 21
The coil of the proportional valve
is defective
Change the proportional valve
AN-903
The proportional valve is
obstructed
Disassemble and clean the
proportional valve.
The flow meter is defective
Change the “Flow meter”
AN-119
The AN-368 PCB is defective
Change the “Bridge PCB AN-
368”.
Summary of Contents for 4
Page 1: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE YB 907...
Page 30: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 30 TECHNICIAN MENU Configuration Adjustments...
Page 31: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 31 SUSPENSION CONFIGURATION...
Page 54: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 54...
Page 58: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 58 LIQUID SENSOR PCB AP 869...
Page 64: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 64 SUCTION DISINFECTION SYSTEM...