XO 4 & XO FLEX TROUBLE SHOOTING GUIDE
Version 3.30
50
DÜRR CAMERA / VIDEO iX, iX HD, HD Smart:
Units manufactured before 08/04/2019 and sold with the iX HD camera are equipped either with the
AO-347
AO-329
“DÜRR Vistacam IX” PCB with external mini USB
connector as shown on page 28.
All units manufactured after 08/04/2019 and sold with the iX HD camera or HD Smart camera, are
equipped with a new
AO-348
“Vistacam iX-HD Smart w. mini USB” block with USB connection.
See picture below
Testcable for Vistacam IX and IX HD with normal bajonet connector as shown on page 28: CA-220.
Testcable for Vistacam IX HD with mini USB in bajonet connector (illustrated below): AP-071 + CA-084.
Failure description
Reason
Solution
The camera does not work
when the suspension is
activated
Suspension cable is not mounted
properly or is defective
Check that the CA-101 cable
connecting the AN-368 PCB to
the AO-329 : iX or AO-347 : iX
HD PCB is properly connected.
Suspension cord is defective
Change the suspension when all
the above has been verified.
Replace the PCB suspension
AO-329 for vistacam iX or
AO-347 for iX HD
Concerns only HD and HD Smart
Cameras
If the camera's LED is always on
while the Unit is off, this indicates
that the AO-348 PCB is defective
The instrument bridge PCB is
defective.
Change Bridge PCB AN-368
Lack of signal strength
Most USB port groups in a
computer share 500 mA power. If
one USB device is connected, it
will get all 500mA.
Summary of Contents for 4
Page 1: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE YB 907...
Page 30: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 30 TECHNICIAN MENU Configuration Adjustments...
Page 31: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 31 SUSPENSION CONFIGURATION...
Page 54: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 54...
Page 58: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 58 LIQUID SENSOR PCB AP 869...
Page 64: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 64 SUCTION DISINFECTION SYSTEM...