XO 4 & XO FLEX TROUBLE SHOOTING GUIDE
Version 3.30
46
SCALER
2 Block
AN-501
3 Water Valve 2/2
AN-506
4 Connector
AN-507
5 PCB
AN-510
6 Cable
CA-101
Configuration
Information
Measurement point
Configuration menu (technician)
- “SUSPx”: Select « Scaler »
- “SPCB”: Select “AD-605”
- « Scaler T »: Select scaler type
(EMS, Satelec, Odontoson).
Failure description
Reason
Solution
The scaler is not activated
when the suspension is
pulled
The suspension cable is not
mounted properly or is defective
Check that the CA-101 cable
connecting the AN-368 PCB to
the suspension PCB is properly
connected.
The suspension cord is
defective.
A little preprogrammed chipset is
placed in the cord.
Change the suspension when all
the above has been verified.
The suspension PCB is
Check if the LED D110 on the