U
SER
M
ANUAL
X
ORAYA
Z7
|
X
ORAYA
ML-Z7
P
RODUCT DESCRIPTION
24
Trigger/wake button (J):
The function of this button depends on the datalogger state. The following
table describes these functions as delivered. Users can configure this
behavior via the X
ORAYA
Suite in the settings of the button interface.
(
➔
State of the datalogger
Function
Sleep mode
Datalogger wakes up
Switched on
Start HDD recording
Switched on, recording active
▪
Short press: Setting a mark (an event) in
the recording
▪
Long press (3 seconds at least): Stop
current recording
Power button (K):
If you press this button while the datalogger is operational, the device
switches to sleep mode.
If the datalogger is operational and this button is pressed together with the
trigger/wake button
(J)
, the datalogger turns off completely.
You cannot turn off the datalogger using this button when the device is
connected to a PC and the
Connect
LED is lit.
In this case, you can turn off the datalogger via the
Logger
menu of the
X
ORAYA
Suite. Here, you may choose between two options, i.e.
Shutdown
and
Shutdown (no wake up)
. (
➔
Power/trigger/wake port (L):
By default, the datalogger must be supplied with 12 V DC voltage.
Optionally, it can also operate with power supplies in a certain specified
range. (
➔
Trigger input and trigger button offer an analogous operation. A +12 V
signal level at the trigger input is equivalent to pressing the trigger button.
(
➔
The wake input allows waking up the datalogger from sleep. To that end,
the wake signal must shift from 0 V to +12 V.
BroadR-Reach LEDs (M):
The housing label
OABR
stands for
OPEN Alliance BroadR-Reach®
.
Port LED is lit green when a connection to a BroadR-Reach component is
established.
Summary of Contents for XORAYA ML-Z7