Turb 430 IR/T
Maintenance, cleaning, disposal
53
ba75507e05
07/2016
5.2
Cleaning
Occasionally wipe the outside of the meter with a damp, lint-free cloth.
Disinfect the housing with isopropanol as required.
CAUTION
The housing components are made out of synthetic materials
(polyurethane, ABS and PMMA). Thus, avoid contact with acetone
and similar detergents that contain solvents. Remove any
splashes immediately.
5.2.1
Cleaning the cell shaft
If liquid is in the cell shaft (e.g. due to a spilled cell), clean the cell shaft
as follows:
5.2.2
Cleaning the cells
Cells have to be clean, dry, and free of fingerprints. Therefore, clean
them regularly:
1
Switch the Turb 430 IR/T off and pull out the power plug.
2
Rinse the cell shaft with distilled water.
1
Clean the cells inside and out with hydrochloric acid or
laboratory soap.
2
Rinse out several times with distilled water.
3
Let them dry in the air.
4
Only hold the cells by the top or by the light protection cap so
that the optical path is not impaired.
5
Before measuring, clean the cell with the enclosed cleaning
cloth.
Summary of Contents for Turb 430 IR/T
Page 1: ...OPERATING MANUAL ba75507e05 07 2016 Turb 430 IR T HANDHELD TURBIDIMETER...
Page 12: ...Safety Turb 430 IR T 12 ba75507e05 07 2016...
Page 16: ...Commissioning Turb 430 IR T 16 ba75507e05 07 2016 LabStation...
Page 18: ...Commissioning Turb 430 IR T 18 ba75507e05 07 2016...
Page 50: ...Operation Turb 430 IR T 50 ba75507e04 06 2009...
Page 56: ...What to do if Turb 430 IR T 56 ba75507e05 07 2016...
Page 60: ...Technical data Turb 430 IR T 60 ba75507e05 07 2016...
Page 66: ...Lists Turb 430 IR T 66 ba75507e05 07 2016...
Page 68: ...Index Turb 430 IR T 68 ba75507e05 07 2016...