Note
Before measuring the third calibration standard of 0.02 FNU/NTU you
can exit the calibration with
<ESC>
at any time.
The new calibration data is discarded. The old calibration data is further
used.
3
Align the cell:
Marked cell:
– Align the marking on the cell cap with the marking on the
cell shaft.
– Press and hold the
<START/ENTER>
key until the
measured value is displayed.
Unmarked cell (see page 33)
– Press and hold the
<START/ENTER>
key.
– Slowly and in small steps turn the cell by one complete
rotation (by 360 °).
– After each step wait for a short time until the displayed
measured value is stable.
– Turn the cell back to the position with the lowest
measured value.
4
Release the
<START/ENTER>
key.
Measurement of the calibration standard begins.
5
Repeat the steps 4 - 6 with the calibration standards, 10.0 FNU/
NTU and 0.02 FNU/NTU.
After measuring the 0.02 FNU/NTU calibration standard, the
calibration result is displayed.
Calibration is completed.
6
Confirm the calibration result with
<START/ENTER>
.
The calibration record is displayed.
7
Confirm the calibration record with
<START/ENTER>
.
The display shows instructions for the first measurement.
i
Turb.
= 1000 FNU/NTU
i
Start calibration
by releasing <START>
Turbidity \ Calibration
Summary of Contents for Turb 430 IR/T
Page 1: ...OPERATING MANUAL ba75507e05 07 2016 Turb 430 IR T HANDHELD TURBIDIMETER...
Page 12: ...Safety Turb 430 IR T 12 ba75507e05 07 2016...
Page 16: ...Commissioning Turb 430 IR T 16 ba75507e05 07 2016 LabStation...
Page 18: ...Commissioning Turb 430 IR T 18 ba75507e05 07 2016...
Page 50: ...Operation Turb 430 IR T 50 ba75507e04 06 2009...
Page 56: ...What to do if Turb 430 IR T 56 ba75507e05 07 2016...
Page 60: ...Technical data Turb 430 IR T 60 ba75507e05 07 2016...
Page 66: ...Lists Turb 430 IR T 66 ba75507e05 07 2016...
Page 68: ...Index Turb 430 IR T 68 ba75507e05 07 2016...