Operation
Turb 430 IR/T
44
ba75507e04
06/2009
4.7.1
Connecting a PC/external printer
Connect the interface to the devices via the AK540/B (PC) or AK540/S
(external printer) cable.
CAUTION
The RS232 interface is not galvanically isolated. When connecting
an earthed PC/printer, measurements cannot be performed in
earthed media as incorrect values would result.
Set up the following transmission data on the PC/printer:
Socket assignment
4.7.2
Configuring the RS232 interface
For error-free data transmission, the RS232 interface should be set to
the same transmission speed (
Baud rate
) on the Turb 430 IR/T and
PC/printer.
You can set the following values for the baud rate on the Turb 430 IR/
T: 1200, 2400, 4800, 9600, 19200.
The baud rate is selected in the menu,
Configuration
/
System
/
Interface
/
Baud rate
.
Baud rate
can be selected from: 1200, 2400, 4800, 9600,
19200
The baud rate must agree with the baud rate set on
the PC/printer.
Handshake
RTS/CTS
PC only:
Parity
none
Data bits
8
Stop bits
1s
RS 232
REC
1
3
4
5
6
2
1 RTS
2 RxD
3 TxD
4 DTR
5 SG
6 CTS
Summary of Contents for Turb 430 IR/T
Page 1: ...OPERATING MANUAL ba75507e05 07 2016 Turb 430 IR T HANDHELD TURBIDIMETER...
Page 12: ...Safety Turb 430 IR T 12 ba75507e05 07 2016...
Page 16: ...Commissioning Turb 430 IR T 16 ba75507e05 07 2016 LabStation...
Page 18: ...Commissioning Turb 430 IR T 18 ba75507e05 07 2016...
Page 50: ...Operation Turb 430 IR T 50 ba75507e04 06 2009...
Page 56: ...What to do if Turb 430 IR T 56 ba75507e05 07 2016...
Page 60: ...Technical data Turb 430 IR T 60 ba75507e05 07 2016...
Page 66: ...Lists Turb 430 IR T 66 ba75507e05 07 2016...
Page 68: ...Index Turb 430 IR T 68 ba75507e05 07 2016...