Turb 430 IR/T
Technical data
59
ba75507e05
07/2016
7.2
Turbidity
7.2.1
Turb 430 IR
7.2.2
Turb 430 T
Measuring principle
Nephelometric measurement according to DIN EN ISO 7027
Light source
Infrared LED
Measuring range
0.01 ... 1100 FNU/NTU
Resolution
in the range
0,01 ... 9,99
max. 0.01 FNU/NTU
in the range
10,0 ... 99,9
max. 0.1 FNU/NTU
in the range
100 ... 1100
max. 1 FNU/NTU
Accuracy
in the range
0 ... 1100 FNU/NTU
± 2 % of the measured value
or ± 0.01 FNU/NTU
Reproducibility
0,5% of the measured value
Response time
4 seconds
Calibration
Automatic 3-point calibration
Measuring principle
Nephelometric measurement according to US EPA 180.1
Light source
White light tungsten lamp
Measuring range
0.01 ... 1100 NTU
Resolution
in the range
0,01 ... 9,99
max. 0.01 NTU
in the range
10,0 ... 99,9
max. 0.1 NTU
in the range
100 ... 1100
max. 1 NTU
Accuracy
in the range
0 ... 500 NTU
± 2 % of the measured value
or ± 0.01 NTU
in the range
500 ... 1100 NTU
± 3 % of the measured value
Reproducibility
1% of the measured value
Response time
7 seconds
Calibration
Automatic 3-point calibration
Summary of Contents for Turb 430 IR/T
Page 1: ...OPERATING MANUAL ba75507e05 07 2016 Turb 430 IR T HANDHELD TURBIDIMETER...
Page 12: ...Safety Turb 430 IR T 12 ba75507e05 07 2016...
Page 16: ...Commissioning Turb 430 IR T 16 ba75507e05 07 2016 LabStation...
Page 18: ...Commissioning Turb 430 IR T 18 ba75507e05 07 2016...
Page 50: ...Operation Turb 430 IR T 50 ba75507e04 06 2009...
Page 56: ...What to do if Turb 430 IR T 56 ba75507e05 07 2016...
Page 60: ...Technical data Turb 430 IR T 60 ba75507e05 07 2016...
Page 66: ...Lists Turb 430 IR T 66 ba75507e05 07 2016...
Page 68: ...Index Turb 430 IR T 68 ba75507e05 07 2016...