Turb 430 IR/T
Operation
35
ba75507e04
06/2009
Note
To keep the drift as low as possible, the time for aligning the cell while
pressing and holding the
<START/ENTER>
key is limited to 30
seconds. After this time, the meter automatically starts measuring or
calibrating.
Display with measuring
range overflow
If the measured value is outside the measuring range of the
Turb 430 IR/T, it is indicated on the display:
6
Align the cell:
Marked cell
– Align the marking on the cell cap with the marking on the
cell shaft.
– Press and for a short time hold the
<START/ENTER>
key until the measured value is displayed.
Unmarked cell (see page 33)
– Press and hold the
<START/ENTER>
key.
– Slowly and in small steps rotate the cell by one complete
rotation (by 360 °). After each step wait for a short time
until the displayed measured value is stable.
– Turn the cell back to the position with the lowest
measured value.
7
Release the
<START/ENTER>
key.
Measurement starts. The measured value is displayed.
8
Repeat the steps 2 to 8 for further samples.
Turbidity
157.0
FNU
NTU
01.02.05 15:12
Turbidity
> 1100
FNU
NTU
01.02.05 15:12
Summary of Contents for Turb 430 IR/T
Page 1: ...OPERATING MANUAL ba75507e05 07 2016 Turb 430 IR T HANDHELD TURBIDIMETER...
Page 12: ...Safety Turb 430 IR T 12 ba75507e05 07 2016...
Page 16: ...Commissioning Turb 430 IR T 16 ba75507e05 07 2016 LabStation...
Page 18: ...Commissioning Turb 430 IR T 18 ba75507e05 07 2016...
Page 50: ...Operation Turb 430 IR T 50 ba75507e04 06 2009...
Page 56: ...What to do if Turb 430 IR T 56 ba75507e05 07 2016...
Page 60: ...Technical data Turb 430 IR T 60 ba75507e05 07 2016...
Page 66: ...Lists Turb 430 IR T 66 ba75507e05 07 2016...
Page 68: ...Index Turb 430 IR T 68 ba75507e05 07 2016...