Turb 430 IR/T
What to do if...
55
ba75507e05
07/2016
6
What to do if...
6.1
General errors
Display,
LoBat
Instrument does not
react to keystroke
Error message,
Error
0, 8, 16, 16384
6.2
Turbidity
Error message,
Measured values
obviously incorrect
Measured value display
< 0.01 FNU
Cause
Remedy
– The batteries or accumulator
pack are largely depleted
– Insert new batteries
– Charge the accumulator pack
(see section 3.2)
Cause
Remedy
– Software error
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the
<START/ENTER>
and
<PRT>
key
simultaneously.
Cause
Remedy
– Instrument error
– Repeat measurement
– Meter defective,
send meter to WTW for repair
and quote the error number
Cause
Remedy
– Cell not correctly inserted
– Lock cell into place
– Cell contaminated
– Clean the cell
– Calibration too old
– Carry out calibration
Cause
Remedy
– Calibration defective
– Carry out calibration
– Measured value outside the
measuring range
– not possible
Summary of Contents for Turb 430 IR/T
Page 1: ...OPERATING MANUAL ba75507e05 07 2016 Turb 430 IR T HANDHELD TURBIDIMETER...
Page 12: ...Safety Turb 430 IR T 12 ba75507e05 07 2016...
Page 16: ...Commissioning Turb 430 IR T 16 ba75507e05 07 2016 LabStation...
Page 18: ...Commissioning Turb 430 IR T 18 ba75507e05 07 2016...
Page 50: ...Operation Turb 430 IR T 50 ba75507e04 06 2009...
Page 56: ...What to do if Turb 430 IR T 56 ba75507e05 07 2016...
Page 60: ...Technical data Turb 430 IR T 60 ba75507e05 07 2016...
Page 66: ...Lists Turb 430 IR T 66 ba75507e05 07 2016...
Page 68: ...Index Turb 430 IR T 68 ba75507e05 07 2016...