4.5
Turbidity
4.5.1
General information
Venting the sample
Air bubbles in the sample affect the measuring result to a massive
extent because they have a large scattering effect on the incident light.
Larger air bubbles cause sudden changes in the measured values
whereas smaller air bubbles are recorded by the instrument as
turbidity. Therefore, avoid or remove air bubbles:
Avoiding or removing
air bubbles
During sampling, ensure all movement is kept to a minimum
If necessary, vent the sample (ultrasonic baths, heating or adding a
surface-active substance to reduce the surface tension)
Note
To measure turbidity values below 1 FNU/NTU, additionally note the
appendix 2 on page 70.
4.5.2
Aligning and marking a cell
Even perfectly clean quality cells exhibit tiny directional differences in
their light transmittance. Therefore, if you want to achieve accurate and
reproducible measurement results, it is necessary to always index the
sample cells and cells for calibration standards in the same way (see
section 2130 of "Standard Methods for the Examination of Water and
Wastewater", 19th edition).
To do so, the optimum alignment of the cell is determined.
Note
Never apply oily liquids (or so-called "special silicone oils") in order to
"smooth" possible scratches. They would unnecessarily soil the meter
and your working environment. The measurement accuracy is ensured
by aligning the cells. Scratched cells have to be replaced.
Aligning the cell
1
Clean the cell (see section 5.2.2).
2
Insert the cell (see section 4.2).
3
Align the cell:
Press and hold the
<START/ENTER>
key.
Slowly and in small steps rotate the cell by one complete
rotation (by 360 °).
After each step wait for a short time until the displayed
measured value is stable.
Turn the cell back to the position with the lowest measured
value.
Summary of Contents for Turb 430 IR/T
Page 1: ...OPERATING MANUAL ba75507e05 07 2016 Turb 430 IR T HANDHELD TURBIDIMETER...
Page 12: ...Safety Turb 430 IR T 12 ba75507e05 07 2016...
Page 16: ...Commissioning Turb 430 IR T 16 ba75507e05 07 2016 LabStation...
Page 18: ...Commissioning Turb 430 IR T 18 ba75507e05 07 2016...
Page 50: ...Operation Turb 430 IR T 50 ba75507e04 06 2009...
Page 56: ...What to do if Turb 430 IR T 56 ba75507e05 07 2016...
Page 60: ...Technical data Turb 430 IR T 60 ba75507e05 07 2016...
Page 66: ...Lists Turb 430 IR T 66 ba75507e05 07 2016...
Page 68: ...Index Turb 430 IR T 68 ba75507e05 07 2016...