Operation
Turb 430 IR/T
36
ba75507e04
06/2009
4.5.4
Settings for turbidity measurements
Overview
For turbidity measurements, the following settings are possible in the
menu,
Configuration
/
Turbidity
:
Calibration record
(display, print)
Entering the
Calibration interval
Reset
Settings/functions
The settings are given in the menu,
Configuration
/
Turbidity
.
Move to the
Configuration
menu with the
<MENU>
key.
4.5.5
Calibration
When to calibrate?
After the calibration interval has expired
With a temperature change
Calibration procedures
and calibration
standards
For the menu-guided three-point calibration you need the following
three calibration standards in the mentioned order:
Menu item
Possible
setting
Description
Calibration record
-
Displays the calibration
record of the last
calibration.
Calibration interval
1 ... 999 d
Calibration interval
for
turbidity measurement (in
days).
If the calibration interval
has expired, the meter
reminds you to calibrate
before each measurement.
Reset
Reset all settings for the
Turbidity
measuring mode
(see section 4.8.2)
Standard no.
FNU/NTU
1
1000
2
10,0
3
0,02
Summary of Contents for Turb 430 IR/T
Page 1: ...OPERATING MANUAL ba75507e05 07 2016 Turb 430 IR T HANDHELD TURBIDIMETER...
Page 12: ...Safety Turb 430 IR T 12 ba75507e05 07 2016...
Page 16: ...Commissioning Turb 430 IR T 16 ba75507e05 07 2016 LabStation...
Page 18: ...Commissioning Turb 430 IR T 18 ba75507e05 07 2016...
Page 50: ...Operation Turb 430 IR T 50 ba75507e04 06 2009...
Page 56: ...What to do if Turb 430 IR T 56 ba75507e05 07 2016...
Page 60: ...Technical data Turb 430 IR T 60 ba75507e05 07 2016...
Page 66: ...Lists Turb 430 IR T 66 ba75507e05 07 2016...
Page 68: ...Index Turb 430 IR T 68 ba75507e05 07 2016...