pHotoFlex / pHotoFlex Turb
What to do if...
91
ba75491e05
04/2009
Measured value display
----
(calibration error)
No stable measured
value
Cause
Remedy
Electrode:
– Diaphragm contaminated
– Clean diaphragm
– Membrane contaminated
– Clean membrane
– Moisture in the plug
– Dry plug
– Not enough electrolyte
– Top up electrolyte
– Electrode obsolete
– Replace electrode
– Electrode broken
– Replace electrode
– Socket damp
– Dry socket
Calibration procedure:
– Incorrect solution temperature
(without temperature sensor)
– Set up correct temperature
– Incorrect buffer solutions
– Select buffer solutions
suitable for the calibration
procedure
– Buffer solutions too old
– Use only once.
Note the shelf life
Cause
Remedy
pH electrode:
– Diaphragm contaminated
– Clean diaphragm
– Membrane contaminated
– Clean membrane
Test sample
– pH value not stable
– Measure with air excluded if
necessary
– Temperature not stable
– Temper if necessary
Ele test sample:
– Conductivity too low
(e.g. in ultrapure water)
– Use suitable electrode
– Temperature too high
– Use suitable electrode
– Organic liquids
– Use suitable electrode
Summary of Contents for pHotoFlex
Page 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Page 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Page 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Page 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Page 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Page 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Page 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Page 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...