Operation
pHotoFlex / pHotoFlex Turb
44
ba75491e05
04/2009
Measuring the blank
value
Note
The following measurement of the blank value applies only to the se-
lected program. If no program is selected, the message
i
No program
selected. appears on the display.
1
Press the <M> key (long pressure) repeatedly until the Pho-
tometry measuring mode is selected.
2
Press the <M> key (short pressure) repeatedly until the mea-
sured variable, Concentration is selected.
3
Select a program with <PROG> as necessary.
4
Open the adjustment menu with <CAL/ZERO>.
5
Using <
▲
> <
▼
> and <START/ENTER>, select and start the
Measure blank value function.
The menu-guided blank value measurement starts.
Follow the instructions on the display.
6
Insert a cell with blank sample (see section 4.2).
7
Start the measurement of the blank value with <START/EN-
TER>.
After measuring, the result of the blank value measurement is
displayed and stored.
The result is displayed as
i
Blank value measurement suc-
cessful! or
i
Blank value measurement erroneous!
Zero adjust.
Measure blank value
Delete blank value
Photometry \ Adjustment
i
2: A5/25 MC
i
Cell = 16 mm
i
Insert blank sample
Start measurement
Measure blank value
Summary of Contents for pHotoFlex
Page 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Page 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Page 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Page 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Page 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Page 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Page 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Page 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...