pHotoFlex / pHotoFlex Turb
110
ba75491e05
04/2009
Appendix 2: Turbidity values under 1 FNU/
NTU
With turbidity values (under 1 FNU/NTU), the measured value is
strongly influenced by the cell and its alignment.
In order to increase measuring accuracy with turbidity values under 1
FNU/NTU, calibration in the 0.02 FNU/NTU standard and later
measurement should be carried out in the same cell. For calibration in
the standards 10.0 and 1000 FNU/NTU follow the instructions on the
display.
Proceed as follows to measure turbidity values under 1 FU/NTU:
Calibration procedure:
Calibrate
z
after the calibration interval has expired
z
with a temperature change
Measuring
1
Press the <CAL/ZERO> key.
The menu-guided calibration begins.
2
Carry out calibration of the standards 1000 FNU/NTU and
10.0 FNU/NTU in a clean and unscratched cell according to
the menu guidance.
3
Fill the cleaned cell with the 0.02 FNU/NTU standard and
calibrate.
4
Mark the alignment of the cell.
5
Fill the marked and cleaned cell with test sample, align it
with the marking and measure.
6
If necessary, fill the marked and cleaned cell once more
with test sample and carry out further measurements.
Summary of Contents for pHotoFlex
Page 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Page 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Page 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Page 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Page 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Page 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Page 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Page 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...