Operation
pHotoFlex / pHotoFlex Turb
66
ba75491e05
04/2009
Display if the
measuring range is
exceeded
If the measured value is outside the measuring range of the
pHotoFlex Turb, it is indicated on the display:
4.7.4
Calibration
When to calibrate?
z
Regularly every 90 days
z
With a temperature change
Calibration procedures
and calibration
standards
For the menu-guided three-point calibration you need the following
three calibration standards in the mentioned order:
Preparing calibration
Perform the following preparatory activities when you want to calibrate:
CAUTION
Never open the cells with the calibration standards.
Turbidity
> 1100
FNU
NTU
01.02.05 15:12
Standard no.
NTU/FNU
1
1000
2
10,0
3
0,02
1
Keep the cells with the required calibration standards ready
and mark them as necessary (see page 62).
2
Clean the cell (see section 5.2.2).
3
Summary of Contents for pHotoFlex
Page 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Page 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Page 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Page 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Page 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Page 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Page 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Page 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...