![wtw pHotoFlex Operating Manual Download Page 67](http://html1.mh-extra.com/html/wtw/photoflex/photoflex_operating-manual_1000666067.webp)
pHotoFlex / pHotoFlex Turb
Operation
67
ba75491e05
04/2009
Carrying out calibration
1
Press the <M> key (long pressure) repeatedly until the Turbid-
ity measuring mode is selected.
2
Press the <CAL/ZERO> key.
The menu-guided calibration begins.
Follow the instructions on the display.
3
Insert the cell with the displayed calibration standard (here e.g.
1000 NTU/FNU) in the cell shaft (see section 4.2).
4
Align the cell:
z
Marked cell:
– Align the marking on the cell cap with the marking on the
cell shaft.
– Press and time hold the <START/ENTER> key until the
measured value is displayed.
z
Unmarked cell (see page 62)
– Press and hold the <START/ENTER> key.
– Slowly and in small steps turn the cell by one complete
rotation (by 360 °).
– After each step wait for a short time until the displayed
measured value is stable.
– Turn the cell back to the position with the lowest mea-
sured value.
5
Release the <START/ENTER> key.
Measurement of the calibration standard begins.
i
Insert standard
1000 FNU/NTU
i
Press and hold <START>
i
Align sample
Turbidity \ Calibration
i
Turb.
= 1000 FNU/NTU
i
Start calibration
by releasing <START>
Turbidity \ Calibration
Summary of Contents for pHotoFlex
Page 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Page 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Page 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Page 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Page 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Page 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Page 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Page 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...