pHotoFlex / pHotoFlex Turb
Operation
61
ba75491e05
04/2009
Continuing for three-
point calibration
(Cal. type TEC)
14
Thoroughly rinse the electrode with distilled water.
15
Immerse the electrode in buffer solution 3.
16
If necessary, measure the temperature of buffer 3 manually,
then enter and confirm it with <
▲
> <
▼
> and <START/ENTER>
in the Set temperature setting.
17
Using <
▲
> <
▼
>, select Continue and press <START/EN-
TER>. The buffer is measured.
The measured value is checked for stability (AutoRead).
18
Wait for the end of the AutoRead measurement or Terminate
AutoRead with <START/ENTER> and take over the calibration
value.
The new calibration values are displayed as an informative
message (
i
).
You have the following options:
z
Accept the new calibration values with <START/ENTER>.
Subsequently, the calibration record is displayed and output
to the interface at the same time.
z
To switch to the measured value display without accepting
the new calibration values, press <M> (short pressure) or
<ESC>.
i
Buffer value = 4.01
i
U = 184 mV
i
Temperature =
24.8 °C
Terminate AutoRead
pH & ORP \ Calibration
Summary of Contents for pHotoFlex
Page 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Page 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Page 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Page 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Page 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Page 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Page 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Page 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...