pHotoFlex / pHotoFlex Turb
Operation
47
ba75491e05
04/2009
Note
If
i
Calibration error! was displayed as the calibration result, a note au-
tomatically reminds you of another zero adjustment before the next
measurement.
Measuring is not possible without a valid zero adjustment.
4
Using <
▲
> <
▼
> and <START/ENTER>, select and start the
Zero adjust. function.
The menu-guided zero adjustment starts.
Follow the instructions on the display.
5
Insert the cell (see section 4.2)
6
Set another cell with <
▲
> <
▼
> and <START/ENTER> as nec-
essary.
7
Start the measurement of the zero adjustment with <START/
ENTER>.
After measuring, the result of the zero adjustment is displayed
and stored.
i
Zero adjust. successful! (successful zero adjustment) or
i
Calibration error! (zero adjustment not successful)
is displayed as the result.
The zero adjustment is completed.
8
Confirm the result with <START/ENTER>.
The zero adjustment is completed.
The meter is ready to measure.
Zero adjust.
Measure blank value
Delete blank value
Photometry \ Adjustment
i
Insert zero
cell (dist. water)
Cell
16 mm
Start measurement
Zero adjust.
Summary of Contents for pHotoFlex
Page 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Page 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Page 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Page 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Page 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Page 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Page 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Page 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...