4.6
pH value / ORP voltage
4.6.1
General information
You can measure the following variables:
z
pH value [ ]
z
ORP [mV]
CAUTION
When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as incorrect values would result. The
RS232 interface is not galvanically isolated.
Temperature
measurement
For reproducible pH measurements, it is essential to measure the tem-
perature of the test sample.
You have the following possibilities for measuring the temperature:
z
Automatic measurement of the temperature by a temperature sen-
sor (NTC30 or Pt1000) integrated in electrode.
z
Manual determination and input of the temperature.
The meter recognizes whether a suitable electrode is connected and
automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature mea-
suring mode:
Preparatory activities
Perform the following preparatory activities when you want to measure:
Temperature
sensor
Resolution of the
temp. display
Temperature of the test
sample
yes
0.1 °C
automatic measurement
-
1 °C
manual measurement
and entry
1
Connect a pH or ORP electrode to the meter.
2
Press the <M> key (long pressure) repeatedly until the pH &
ORP measuring mode is selected.
3
Press the <M> key (short pressure) repeatedly until the mea-
sured parameter, pH or ORP is selected.
4
Adjust the temperature of the solutions and measure the cur-
rent temperature if the measurement is made without a temper-
ature sensor.
5
Calibrate or check the meter with the electrode.
Summary of Contents for pHotoFlex
Page 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Page 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Page 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Page 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Page 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Page 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Page 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Page 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...